Model F5K-10-02 High Voltage Pulser

5,000 volts - 10 amp Maximum Current

Variable Width

The Model F5k-10-02 is a general purpose laboratory instrument which operates under dynamic load conditions. The maximum current of 10 amps permits operating with load resistances down to 500 ohms or with high impedance loads which require a parallel "pull down" resistor to decrease fall time. The instrument consists of a single planar triode switch with pulse fall time a function of load RC. The pulser is intended for high pulse current or fast rise time applications. For low pulse current applications see the M Series pulsers.

Applications

    Physics/Chemistry Research
    Semiconductor Materials Research
    Electric Fields in Aqueous Solutions
    High Voltage Testing
    Liquid Cells

Operating Features

    Power Supply Digital Panel Meter
    Local/Remote Operation with Electrical Interface
    Front Panel HV On Control and Indicator
    Front Panel HV Off Control and Indicator
    Front Panel Pulse Amplitude Adjustment
    Integral Pulse Voltage and Current Monitor
    System Status LED's

Safety Features

    Zero Restart for High Voltage On (requires voltage to be set to zero before restart)
    Auto Shut-down with Excessive Average Power Supply Current
    Auto Capacitance Discharge when High Voltage Off
    External Interlock

Pulser Performance Specifications

The following specifications represent typical values at 10 amps, 7,000 volt power supply, specifications at 5,000 volts:

Parameter Units 10 amps
Amplitude volts 100 to 5,000
Polarity negative
Maximum Peak Current amps >10
Pulse Width (at 50%) µs 0.2 to 100
Rise Time (10% to 90%) ns <50
Fall Time ns <120
Droop <5%
Max Rep Rate Hertz >5,000
Duty Cycle >0.06%

Figure 1 - Load Resistance (ohms) vs Pulse Amplitude (volts)
Figure 2 - Load Resistance (ohms) vs Fall Time (ns)
Figure 3 - Pulse Width (µs) vs Continuous Operation Life (hours)

Front Panel Functions

    Line Power On/Off with Indicator
    High Voltage Controls (Local Mode) Status

Back Panel Functions

    Line Power In - IEC Connector and EMI Filter
    Fuse (3AG, 2 amp, SLO BLO)
    External Interlock - 2 position Terminal Block
    High Voltage Out - SHV Connector
    Pulse Current Monitor - BNC (0.5 v/a into 50 ohms)
    Pulse Voltage Monitor Out - BNC (1000:1, into 50 ohms)
    System Ground Bolt - 1/4-20 Bolt and Nut
    Local Remote Toggle Switch

Electrical Interface - DB9
Pin 1 Remote HV On/Off (On = +2.5 to +12 vdc)
Pin 2 Remote HV Supply Program (0 to 5 vdc)
Pin 3 HV Supply Enable Out (Disabled = 0 vdc, Enabled = 5vdc)
Pin 4 Power Supply Average Current Monitor
Pin 5 Ground
Pin 6 N/C
Pin 7 N/C
Pin 8 Power Supply Voltage Monitor
Pin 9 Pulse Enabled
Mechanical and Line Power
Line In 100 to 120 vac, 50/60 Hz, 1 amp
Size 7 in h x 17 in w x 17 in d (18 cm h x 43 cm w x 43 cm d)
Weight 40 pounds (18 kg)
Operating Temperature 10 to 40°C
Options
Line Power 200 to 240 vac, 50/60 Hz
Polarity positive
Rack Mount
High Voltage Supply 60 Watt

Warranty Information

HV Pulse products are warranted against defects in materials and workmanship. This limited warranty is valid for one year from the date of shipment.